Preparation of Measurement Accessories

Connecting the Accessories

Connecting the Accessories
Step 1. Connect the measurement accessories to the Agilent 4294A.
Step 2. Press the power switch on the front panel of the Agilent 4294A to turn it on.
For more information on connecting accessories to the Agilent 4294A for measurement,
refer to the operation manual furnished with each accessory.
NOTE
Whenever you connect accessories to or disconnect accessories from the Agilent 4294A,
you must first turn off the dc bias (Key strokes:
this step is not taken, the dc bias may destroy the device.
When you connect or disconnect accessories with the Agilent 4294A turned on, error
messages such as "ADC OVERLOAD" or "BRIDGE UNBALANCED" may be displayed
on the LCD. However, this does not cause any special problem. To turn off the message,
press
The power-on self-test is executed at power-on. If an error is detected in the power-on
self-test, the message "CAUTION: POWER ON TEST FAILED" is displayed in the
upper-left area of the screen. In this case, contact your nearest Agilent Technologies office
or the company from which you purchased this instrument. For more information on the
self-test (self-diagnostic) function, refer to "Performing Self-Diagnosis of the Agilent
4294A" on page 320.
66
key.
[Entry Off]
-
-
[Source]
BIAS MENU
BIAS ON off
). If
Chapter 4