Preparation of Measurement Accessories
Adapter Setting
Figure 4-5
Connecting OPEN (0 S) to the 42942A
Step 6. Press the
NOTE
The traces on the LCD display do not change during the phase compensation data
measurement.
When the phase compensation data measurement is completed, the softkey label changes
to
Step 7. With the OPEN (0 S) standard connected to the 7-mm port, press the
the open data measurement.
NOTE
When the load data measurement is completed, the softkey label changes to
OPEN [DONE]
Step 8. Remove the OPEN (0 S) standard from the 7-mm port of the 42942A. Then connect the
SHORT (0 W) standard to the 7-mm port (the SHORT standard is furnished with the
42942A Terminal Adapter).
76
key to start the phase compensation data measurement.
PHASE COMP [-]
.
PHASE COMP [DONE]
.
key to start
OPEN [-]