Table B-47
List sweep (List Sweep) menu
Key operation:
[Sweep] - EDIT LIST
Softkey
PEN
Sets the pen number for the trace of the list sweep segment specified by the
pointer (>). There are six graphics pens that have different colors. See Table
B-24, "Item color modification (Color Item) menu (2/3)," on page 371, or Table
B-25, "Item color modification (Color Item) menu (3/3)," on page 371 to
modify each pen color.
CLEAR LIST ä
Displays Table B-52, "List deletion (Clear List?) menu," on page 387.
done
Stores the edited list sweep table, and returns to Table B-43, "Sweep (Sweep)
menu," on page 383.
cancel
Cancels editing of the list sweep table and returns to Table B-43, "Sweep
(Sweep) menu," on page 383.
Table B-48
Segment edit (Edit Segment) menu (1/3)
Key operation:
[Sweep] - EDIT LIST - EDIT
[Sweep] - EDIT LIST - ADD
Softkey
START
Sets the sweep start frequency of the list sweep segment
STOP
Sets the sweep stop frequency of the list sweep segment
CENTER
Sets the sweep center frequency of the list sweep segment
SPAN
Sets the sweep frequency span of the list sweep segment
NUMBER OF POINTS
Sets the number of points in the list sweep segment. Specify a value between 2
and 201 as the number of sweep points per segment. Note that the maximum
number of points in all segments is a total of 801.
more 1/3
Displays Table B-49, "Segment edit (Edit Segment) menu (2/3)," on page 385.
done
Stores the edited list sweep segment and returns to Table B-47, "List sweep (List
Sweep) menu," on page 384.
cancel
Cancels the editing of the list sweep segment and returns to Table B-47, "List
sweep (List Sweep) menu," on page 384.
1. The frequency range specified for each segment may overlap the frequency ranges of other segments.
Table B-49
Segment edit (Edit Segment) menu (2/3)
Key operation:
[Sweep] - EDIT LIST - EDIT - more 1/3
[Sweep] - EDIT LIST - ADD - more 1/3
Softkey
OSC MODE [ ]
Switches the oscillator level of the list sweep segment between the voltage
source mode [VOLT] and the current source mode [CURR].
OSC LEVEL
Sets the oscillator level of the list sweep segment.
Appendix B
Softkeys displayed by pressing the [Sweep] key
Function
Function
1
.
1
.
1
.
1
.
Function
Key Definitions
GPIB command
SCOL
CLEL
EDITDONE
(N/A)
GPIB command
STAR <value>
STOP <value>
CENT <value>
SPAN <value>
POIN <value>
(N/A)
SDON
SQUI
GPIB command
POWMOD
{VOLT|CURR}
POWE <value>
385