Step 2. Activate the trace for testing the limit line.
Active Trace
Trace A
Trace B
Step 3. Press the
NOTE
The limit line table can be set without displaying the marker. However, the presence of a
marker will be convenient for your work because it allows the numerical value on the trace
to be read and the marker position to be used to enter the limit line set value.
Step 4. Press the
Step 5. Press the
Step 6. Press the
Limit Line Edit menu.
Figure 8-24
Limit line edit mode
Step 7. Press the
Segment Edit menu. The initial value is input in advance to the segment added to the limit
line table.
Step 8. Press the
area of the screen.
Step 9. Use the keys or rotary knob of the ENTRY block in one of the following ways to specify
the segment start value.
Enter the desired value with the numeric keys (
of the unit keys (
Turn the rotary knob (
Chapter 8
Set a limit to the trace and make pass/fail evaluation
key to display the main marker on the trace.
[Marker]
key to display the System menu.
[System]
key to display the Limit Test menu.
LIMIT TEST
key to invoke the limit line edit mode. This also displays the
EDIT LIMIT LINE
key and add a new segment to the limit line table. This also displays the
ADD
key. The current start value of the segment is displayed in the upper left
START
,
,
[G/n]
[M/m]
[k/m]
) until the desired value is set.
m m m m
Analysis and Processing of Result
Key operation
[A]
[B]
to
,
and
[0]
[9]
[-]
[.]
, or
).
[´1]
) and then press one
257