Reading the difference from the reference point on the screen (delta marker)

Reading the difference from the reference point on the screen
(delta marker)
By placing the marker used as the reference at any point on the screen, you can set and
display the positions (sweep parameter value and measurement parameter value) of the
markers (main marker and sub-marker) based on relative values (differences) from the
reference point. This marker used as the reference is called the delta marker. You can
determine whether it will be placed on the trace (delta-marker-on-the-trace mode) or at any
point on the screen (fixed delta marker mode).
There are two ways to place (move) the delta marker at a reference point on the screen:
using the main marker or using the delta marker alone.

Placing the delta marker on the reference point with the main marker

Follow these steps to first move the main marker to a reference point and then place the
delta marker at that point.
Step 1. Activate the trace that you want to analyze.
Active trace
Trace A
Trace B
Step 2. Using an appropriate function, such as the search function, move the main marker to the
point on the trace that you want to specify as the reference. For how to move the main
marker, refer to related topics including:
"Specify the sweep parameter value and read the value on the trace" on page 212
"Search the maximum/minimum measurements" on page 223
"Search the point of target measurement" on page 225
"Search the maximum/minimum peak" on page 228
Step 3. Press the
key to display the Marker menu.
[Marker]
Step 4. Press the
DMODE MENU
Step 5. Select the delta-marker-on-the-trace mode or the fixed delta marker mode.
Delta marker mode
Delta-marker-on-the-trace mode
(when the trace changes, the delta
marker moves along with it; you set
its position with the sweep parameter
value only)
Chapter 8
Analysis and Processing of Result
Key operation
[A]
[B]
key to display the Delta Marker menu.
Key operation
D
MKR
217