Agilent Technologies 4294A Operating Manual page 259

Precision impedance analyzer
Hide thumbs Also See for 4294A:
Table of Contents

Advertisement

Table 8-3
Selection of equivalent circuit
Equivalent circuit model
E
1. Measurement parameter: | Z | - q, Sweep type: log, Vertical axis: | Z | is log and q is linear
2. Measurement parameter: | Z | - q, Sweep type: linear (or log), Vertical axis: | Z | is log and q is linear
Step 8. Press the
circuit parameter.
When the equivalent circuit parameter is on, the calculated equivalent circuit parameter
will be shown under the equivalent circuit diagram. In the case of measurement trace
display (equivalent circuit diagram is off), the calculated equivalent circuit parameter will
be shown in the equivalent circuit parameter area on the left of the screen.
Figure 8-20
Result of equivalent circuit parameter calculation (using equivalent circuit B)
NOTE
When the partial search range is set, the equivalent circuit parameter is calculated based on
the trace within that range. For the partial search range setting procedure, see "Specify
partial search range" on page 267.
The display of the equivalent circuit parameter on the measurement trace screen can be
turned on and off with the
equivalent circuit parameter has been calculated by pressing the
CALCULATE PARAMETER
Chapter 8
Calculate the equivalent circuit parameter and simulate the frequency
Typical frequency characteristics
CALCULATE PARAMETERS
PARAMETERS ON off
key, the parameter indication is turned on automatically.
Analysis and Processing of Result
Sample (example)
2
Resonator
(or
) key to calculate the equivalent
CALC PARMS
key (
PARAMETERS on OFF
characteristics
). When the
251