F. Calibration When a Probe Adapter is Connected to the
42941A
Perform the following calibration when a probe adapter supplied with the 42941A
Impedance Probe is directly connected to the 3.5-mm port of the 42941A, which is
connected to the four-terminal pair port of the Agilent 4294A, and a device is connected to
the probe adapter.
Figure 6-6

Calibration When a Probe Adapter is Connected to the 42941A

This calibration procedure is applied when the following probe adapters are connected to
the 3.5-mm port of the 42941A (see the operation manual of the 42941A for details of
probe adapters):
Pin Probe (with ground electrode or ground lead)
BNC(m) Adapter
BNC(m) Adapter with alligator clip leads
Step 1. Make sure that the adapter selection is set to "42941A Impedance Probe
(
PROBE 42941A
Chapter 6
F. Calibration When a Probe Adapter is Connected to the 42941A
)."
Calibration
143