Turning The Fixture Compensation On Or Off; Defining The Standard Values For Fixture Compensation - Agilent Technologies 4294A Operating Manual

Precision impedance analyzer
Hide thumbs Also See for 4294A:
Table of Contents

Advertisement

Turning the fixture compensation on or off

The fixture compensation function is automatically turned on after completing each
compensation data measurement. However, you can turn the function on or off as required.
Step 1. Press the
Step 2. Press the
Step 3. Press the
LOAD on OFF
invalid (off) for measurement (pressing these keys toggles your selections).
Softkey Label
OPEN on OFF
OPEN ON off
Softkey Label
SHORT on OFF
SHORT ON off
Softkey Label
LOAD on OFF
LOAD ON off

Defining the standard values for fixture compensation

You can define standard values of OPEN, SHORT, and LOAD for fixture compensation.
Figure 6-9 shows the circuit models of the fixture compensation kit used for the Agilent
4294A.
Figure 6-9
Circuit Models of Fixture Compensation Kit
Each compensation standard value can be confirmed and changed as follows:
Step 1. Press the
Step 2. Press the
Step 3. Press the
Chapter 6
key to display the Calibration Menu.
[Cal]
key to bring up the Fixture Compensation Menu.
FIXTURE COMPEN
(
OPEN on OFF
OPEN ON off
(
) key to make each fixture compensation valid (on) or
LOAD ON off
key to display the Calibration Menu.
[Cal]
key to display the Fixture Compensation Menu.
FIXTURE COMPEN
key to display the Compensation Data Definition Menu.
DEFINE VALUE
Fixture Compensation
) key,
(
SHORT on OFF
SHORT ON off
Fixture OPEN Compensation
Off
On
Fixture SHORT Compensation
Off
On
Fixture LOAD Compensation
Off
On
Calibration
) key, and
153