Softkeys displayed by pressing the
Table B-11
Format (Format) menu (1)
Key operation:
[Format]
Softkey
LIN
Selects the linear scale as the vertical axis (Y axis) for the graph.
|
LOG
Selects the log (logarithm) scale as the vertical axis (Y axis) for the graph.
PHASE UNIT [ ]
Toggles the phase display unit between degree [DEG] and radian [RAD] each
time this key is pressed.
EXP PHASE on OFF
Toggles on and off the phase expansion display function each time this key is
pressed. ON off in the softkey label indicates ON state; on OFF, OFF state.
1. When the measurement parameter is not COMPLEX Z-Y.
Table B-12
Format (Format) menu (2)
Key operation:
[Format]
Softkey
COMPLEX
Selects the complex plane to display the graph. This plane has a grid scale whose
|
horizontal axis is the actual axis and whose vertical axis is an imaginary axis.
POLAR
Selects the polar coordinates to display the graph. Complex parameters are
displayed in an absolute value scale (distance from the origin) and a phase scale
(counterclockwise phase angle).
PHASE UNIT [ ]
Toggles the phase display unit between degree [DEG] and radian [RAD] each
time this key is pressed.
EXP PHASE on OFF
Toggles on and off the phase expansion display function each time this key is
pressed. ON off in the softkey label indicates ON state; on OFF, OFF state.
1. When the measurement parameter is COMPLEX Z-Y. (Key operation: [Meas] - more 1/3 - more 2/3 - COMPLEX Z-Y)
Appendix B

Softkeys displayed by pressing the [Format] key

1
Function
1
Function
Key Definitions
key
[Format]
GPIB command
FMT LINY
FMT LOGY
PHAU {DEG|RAD}
EXPP {ON|OFF}
GPIB command
FMT COMP
FMT POLA
PHAU {DEG|RAD}
EXPP {ON|OFF}
365