Setting Measurement Conditions
Setting and Applying dc Bias
selected is displayed in the key label
Softkey label
MEAS RANGE [1mA]
MEAS RANGE [10mA]
MEAS RANGE [100mA]
NOTE
If the message "CAUTION: DCBIAS OVERLOAD" (error number: 137) still appears
even when the dc bias range is set to 100 mA, the connected DUT requires a dc bias with a
current that exceeds the maximum dc bias current (100 mA) provided by the Agilent
4294A. In this case, decrease the dc bias voltage level to avoid the error.
Figure 5-15
Characteristics of Capacitance of Varactor Diode versus dc Bias Voltage
110
of the
[ ]
MEAS RANGE [ ]
Selected dc bias range
1 mA
10 mA
100 mA
key.
Chapter 5