194
EEPROM WRITE FAIL
A write error is detected in the memory (EEPROM) to store the OPEN, SHORT, and
LOAD data for the four-terminal pair extension setting. Contact your local Agilent
Technologies sales office or the company you purchased this instrument from.
-200
Execution error
A comprehensive execution error occurs and the Agilent 4294A cannot detect further
details of the error. This error code simply indicates the occurrence of an execution error
defined in IEEE 488.2, 11.5.1.1.5.
-123
Exponent too large
The absolute value of the exponent exceeds 32,000. (Refer to IEEE 488.2, 7.7.2.4.1.)
-178
Expression data not allowed
An equation data element was received at where the 4294A did not accept any equation
data elements.
-170
Expression error
An error not described by error numbers from -171 to -179 occurs during the syntax
analysis of equation data.
-257
FILE NAME ERROR
There is an error in the filename and, as a result, the command is not executed correctly.
This message is displayed, for example, when you attempt to copy a file using the same
filename.
-256
File name not found
The specified filename is not found and, as a result, the command is not executed correctly.
This message is displayed, for example, when you attempt to read/write a file that does not
exist on the disk.
143
FLOATING POINT ERROR OCCURRED
Due to the execution of an application inside the instrument for inappropriate data sent
from outside the instrument, an error for floating point operations occurs inside the
instrument. The result of the operation may be incorrect. Contact your local Agilent
Technologies sales office or the company you purchased this instrument from.
85
FORMAT NOT VALID FOR COMPLEX MEASUREMENT
When COMPLEX Z-Y (i.e., complex parameter measurement) is selected as a
measurement parameter, you attempt to select a display format not available for the
measurement parameter, for example, the linear scale format (FMT LINY command), and,
as a result, the command is ignored. When COMPLEX Z-Y is selected as a measurement
parameter, set the display format to the complex plane display format (command: FMT
COMP) or the polar coordinates display format (command: FMT POLA).
86
FORMAT NOT VALID FOR SCALAR MEASUREMENT
Appendix C
EEPROM WRITE FAIL
E
F
Error messages
435