Power Cable Receptacle (To Line); High Stability Frequency Reference Output Connector (Ref Oven, Option 1E5 Only); External Reference Signal Input Connector (Ref In) - Agilent Technologies ENA Series User Manual

Rf network analyzers
Hide thumbs Also See for ENA Series:
Table of Contents

Advertisement

Overview of Functions
Rear Panel: Names and Functions of Parts

8. Power Cable Receptacle (to LINE)

The receptacle (outlet) to which the power cable is connected.
NOTE
To connect the device to a power source (outlet), use the supplied three-prong power cable
with a ground conductor.
The plug attached to the power cable (on the power outlet side or device side of the cable)
serves as the disconnecting device (device that cuts off power supply) of the
E5070B/E5071B. When the power supply must be cut off to avoid such danger as electric
shock, pull out the power cable plug (on the power outlet side or device side of the cable).
For the procedure for turning off the mains in normal use, see the description in "1.
Standby Switch" on page 31.
For more on the power supply, see Chapter 2 "Installation" in the Installation and Quick
Start Guide.
9. High Stability Frequency Reference Output Connector (Ref Oven,
Option 1E5 only)
When Option 1E5 (high stability frequency reference) is installed, the reference signal is
output from this connector.
Connector type: BNC connector, female
Output signal (Nominal): 10 MHz, +0 dBm
NOTE
When Option 1E5 (high stability frequency reference) is installed, connect this connector
to the "10. External Reference Signal Input Connector (Ref In)" on page 56 by using the
BNC(m)-BNC(m) cable included with the option.

10. External Reference Signal Input Connector (Ref In)

The reference signal input connector for phase-locking the measurement signal from the
E5070B/E5071B to the external frequency reference signal. Inputting the reference signal
to this connector improves the accuracy and frequency stability of the measurement signal
from the E5070B/E5071B.
Connector type: BNC connector, female
Input signal (Nominal): 10 MHz ±10 ppm, +0 dBm
NOTE
When the frequency reference signal is input to this connector, the measurement signal
from the E5070B/E5071B is automatically phase-locked to the reference signal. When an
input signal is not present, the frequency reference signal inside the E5070B/E5071B is
automatically used. The
system is phase-locked to the external reference signal and in gray when not phase-locked.
When using Option 1E5 (high stability frequency reference), connect this connector to the

"9. High Stability Frequency Reference Output Connector (Ref Oven, Option 1E5 only)"

on page 56 by using the BNC(m)-BNC(m) cable included with the option.
56
±3 dB
on the instrument status bar is displayed in blue when the
ExtRef
±3 dB