13.3 Storing the battery

Requirement
1. Insert the battery in the battery compartment and store the
Material damage due to prolonged storage of the battery
without recharging!
Storing the battery for a prolonged period of time without
recharging can result in the rapid shutdown of and irreparable
damage to the battery.
⇒ When the battery is stored in the device without a power
⇒ If the battery is not stored in the device: Charge battery every
2. Charge battery at regular intervals:
Type of storage
In device without a power supply
Outside the device
Result
The battery is stored in a dry place and is ready for use.
The device and the battery have been cleaned and disinfected
(see "9.3 Hygienic reprocessing of the device", page
The battery is fully charged.
If available: The replacement battery is fully charged.
device in a dry place.
or
Store the battery in a dry place outside of the device.
supply: Charge battery every 3 months.
5 months.
Charging interval
Every 3 months
Every 5 months