Storing The Battery - Weinmann MEDUMAT Standard2 Instructions For Use Manual

Ventilator
Hide thumbs Also See for MEDUMAT Standard2:
Table of Contents

Advertisement

12 Storage

12.3 Storing the battery

Requirement
Result
212
EN
MEDUMAT Standard
The device and the battery have been cleaned and disinfected
(see "8.3 Hygienic reprocessing of the device",
The battery is fully charged.
If available: The replacement battery is fully charged.
1. Insert the battery in the battery compartment and store the
device in a dry place.
or
Store the battery in a dry place outside of the device.
Material damage due to prolonged storage of the battery
without recharging!
Storing the battery for a prolonged period of time without
recharging can result in the rapid shutdown of and irreparable
damage to the battery.
 When the battery is stored in the device without a power
connection: Charge battery every 3 months.
 If the battery is not stored in the device: Charge battery every
5 months.
2. Charge battery at regular intervals:
Type of storage
In device without power connection
Not in device
The battery is stored in a dry place and is ready for use.
2
page 165).
Charging interval
Every 3 months
Every 5 months

Advertisement

Table of Contents
loading

This manual is also suitable for:

Wm 29500

Table of Contents