Maintenance; Intervals; Check The Battery/Rechargeable Battery.18 - Weinmann Homecare MEDUCORE Easy Service And Repair Instructions

Automatic external defibrillator. device nos. 1000 to 1340 from device no. 1341 with ilcor 2005
Table of Contents

Advertisement

5. Maintenance

5.1 Intervals

5.2 Check the Battery/Rechargeable Battery.
Conditions
18
Maintenance
Note:
After all maintenance and repair work the device must be
tested, as described in chapter „4. Checking the
device"on page 9.
As a preventative maintenance measure
the device must be fully checked:
The points of maintenance are as follows:
Testing for completeness (see chapter 4.10,
page 14)
Visual test (see chapter 4.10, page 14)
– Mechanical damage
– Electrode connections
– Labelling
Check the charge levels on the Battery/
Rechargeable Battery (see chapter 5.2, page 18)
Replace internal battery (see chapter 8.8,
page 29)
Final check in accordance with test directive
WM 40006 (see chapter 4., page 9)
Renew maintenance sticker
a safety related check (SRC) must be
In Germany
conducted within a statutory limit of 2 years in
accordance with §6 Medicinal Products and Users
Ordinance.
The Programme EasyView must be started up and
the PC infrared interface or the infrared adapter
must be aligned with the infrared interface of the
MEDUCORE Easy (see chapter "Operation" in
the PC-software operating instructions EasyView
WM 16880).
every 6 years

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Homecare MEDUCORE Easy and is the answer not in the manual?

This manual is also suitable for:

Wm 40000Wm 40005

Table of Contents