9.5 Fittings

Warning!
Risk of explosion! Never immerse the fittings in disinfectant or other liquids. Only
carry out wipe disinfection. Liquid must not be allowed to enter the pressure reducer.
Otherwise there is a risk of explosion.
If it is absolutely necessary to clean the fittings (e.g., pressure reducer, valve), use a clean
cloth. The cloth may be dry or moistened with clean water.

9.6 Hygiene input filter (optional)

Before changing the hygiene input filter, wipe-disinfect the outside of the unit with the hy-
giene input filter installed.
To guarantee filter efficiency, the hygiene input filter has a life of 24 hours of ventilation
(at FiO
<
100% oxygen) or a maximum of 6 months.
2
The hygiene input filter is a single-use device. Do not reprocess it. Dispose of it correctly.

9.7 Cleaning, disinfection and sterilization

Warning!
Risk of explosion! If wipe disinfection is required, make sure that no liquid enters the
connections. Cleaning agents containing alcohol or grease become flammable when
combined with compressed oxygen and can cause explosions.
Carry out hygienic preparation of the MEDUMAT Transport and the accessories used, as
described in the following table.
Refer to the instructions supplied with the hose system and with the disinfectant used. We
recommend using gigasept FF
wipe disinfection. You are advised to use suitable gloves for disinfection work (e.g., house-
hold or disposable gloves).
Notice:
The minimum service life of reusable components is at least
30 preparation cycles.
You can find further information about hygienic preparation and a list of
all the suitable cleaning agents and disinfectants in a brochure on the
Internet at www.weinmann-emergency.com.
®
(new) for immersion disinfection and terralin
Hygienic preparation
®
protect for
EN