8.7 Cleaning and disinfection plan

Level of
Part
disinfec-
tion
Portable unit and wall mounting
Portable unit
LIFE-BASE
Intermediate
Level
Wall moun-
Disinfection
ting BASE-
STATION
Level of
disinfection
High Level
Disinfection
Intermediate Level
Disinfection
Low Level
Disinfection
Carry out hygienic reprocessing according to the following table
after every use:
Manual
Wipe disin-
cleaning
fection
Wipe down
with
®
neodisher
Wipe down
MediClean
all surfaces
forte
once with
(Dr. Weigert).
Incidin™
Dosage:
OxyWipe S
10 ml/l.
(Ecolab).
Treat all
Exposure
surfaces at
time: 5 min
least twice
until they are
visibly clean
8 Hygienic reprocessing
Description
A germ-killing process with a sterilizing agent
under non-sterile conditions. The process kills
all types of microbial life with the exception of
a large number of bacterial spores.
A germ-killing process using a disinfectant that
kills viruses, mycobacteria, fungi and vegetative
bacteria, but not bacterial spores.
A germ-killing process using a disinfectant that
kills some fungi, vegetative types of bacteria
and lipid viruses.
Immersion
Mechani-
disinfec-
cal repro-
tion
cessing
Not permitted Not permitted Not permitted
LIFE-BASE
Steriliza-
tion
EN
155