3. Undo the two mounting screws and remove
3/2 solenoid valve
4. Replace the seal. Ensure that the seal is posi-
tioned correctly.
5. Screw new 3/2 solenoid valve
6. Refit the pneumatic block (see "6.15 Replac-
ing the pneumatic block with angled outlet" on
page 34).
7. Close the device (see "6.5 Closing the
device" on page 23).
8. Perform a functional check (see "3. Final
Check" on page 8).

6.17 Changing the Air Mix/No Air Mix switch

Tools required:
Crosstip screwdriver, size 2,
L-handled socket wrench 10 mm,
Open-ended spanner SW 17,
Special tool WM 22829 from special tool set WM 15349,
If necessary, side nippers,
If necessary, cable tie,
Vice with protective jaws,
Calibration device WM 22836.
1.
Up to appliance no. 1019 (Basic) or 1399
(Basic p):
Remove the pneumatic block from the housing
(see „6.14 Replacing the pneumatic block" on
page 32, steps
to
1.
From appliance no. 1020 (Basic) or 1400
(Basic p):
Remove the pneumatic block from the housing
(see „6.15 Replacing the pneumatic block with
angled outlet" on page 34, steps
2. Clamp the pneumatic block in a vice with pro-
tective jaws.
3. Unscrew the rocker using an open-ended
spanner (SW 17).
4. Screw in a new rocker
Take care to ensure the correct installation
position:
The rocker must drop automatically into its end
position. It must not become stuck in an interme-
diate position.