Appendix A: Maintenance
Figure 111: Fuse replacement
LINE FUSE
SLOWBLOW
5.0 A, 250 V
LINE RATING
100-240 VAC
50, 60 Hz
500VA MAX.
Perform the following steps to replace the line fuse:
1. Power off the unit and remove the line cord.
2. The fuse drawer (1) is located below the AC receptacle. A small tab is located on the top of the
fuse drawer (2). Using a thin-bladed knife or a screwdriver, pry this tab away from the AC
receptacle.
3. Slide the fuse drawer out to gain access to the fuse (the fuse drawer does not pull completely out
of the power module).
4. Snap the fuse out of the drawer and replace it with the same type (the fuse is specified in the
table below).
To prevent instrument damage, use only the correct fuse type (see table).
5. Push the fuse drawer back into the module.
If the power line fuse continues to blow, a circuit malfunction exists and must be corrected. Return the
unit to Keithley Instruments for repair.
Line fuse
Line voltage
100 V - 240 V

Front panel tests

There are two front panel tests: one to test the functionality of the front panel keys and one to test the
display.
A-2
Model 2651A High Power System SourceMeter® Instrument Reference Manual
DIGITAL I/O
RS-232
IEEE-488
!
!
LAN
2
Rating
5 A, 250 V, FAST-
ACTING CARTRIDGE
FUSE, 5 × 20mm
SENSE/GUARD
OUTPUT
S
S
LO
HI
LO
G
G G G
HI
40V, 50A MAX.
CHANNEL A
!
1
CAUTION
Keithley Instruments
part number
FU-154-5
2651A-901-01 Rev. A / March 2011