Status.operation.remote - Keithley 2651A Reference Manual

High power system sourcemeter instrument
Hide thumbs Also See for 2651A:
Table of Contents

Advertisement

Model 2651A High Power System SourceMeter® Instrument Reference Manual
Details
These attributes are used to read or write to the operation status measuring summary registers. Reading a status
register returns a value. The binary equivalent of the returned value indicates which register bits are set. The
least significant bit of the binary number is Bit B0, and the most significant bit is Bit B15.
For information about .condition, .enable, .event, .ntr, and .ptr registers, refer to
2) and
Enable and transition registers
meanings:
Bit
Value
B0
Not used
B1
status.operation.measuring.SMUA
B2-B15
Not used
In addition to the above constant, operationRegister can be set to the decimal equivalent of the bit to set.
Example
status.operation.measuring.enable =
status.operation.measuring.SMUA
Also see
Operation Status Registers
status.operation.*

status.operation.remote.*

This attribute contains the operation status remote summary register set.
Type
TSP-Link accessible
Attribute
- -
.condition (R)
Yes
.enable (RW)
Yes
.event (R)
Yes
.ntr (RW)
Yes
.ptr (RW)
Yes
2651A-901-01 Rev. A / March 2011
(on page E-18). The individual bits of this register have the following
(on page E-20)
(on page 7-266)
Affected by
- -
Not applicable
Status reset
Status reset
Status reset
Status reset
Status register sets
Description
Not applicable.
Bit will be set when taking an overlapped
measurement, but it will not set when taking a
normal synchronous measurement.
Bit B1 decimal value: 2
Binary value: 0000 0010
Not applicable.
Sets the SMUA bit of the operation
status measuring summary enable
register using a constant.
Where saved
Default value
- -
- -
Not saved
Not applicable
Not saved
0
Not saved
0
Not saved
0
Not saved
2050 (All bits set)
Section 7: Command reference
(on page E-
7-293

Advertisement

Table of Contents
loading