Model 2651A High Power System SourceMeter® Instrument Reference Manual

smuX.source.sink

This attribute turns sink mode on or off.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
sinkMode = smuX.source.sink
smuX.source.sink = sinkMode
sinkMode
X
Details
This attribute enables or disables sink mode. When enabled, sink mode reduces the source limit inaccuracy seen
when operating in quadrants II and IV (quadrants I and III will show this source limit inaccuracy).
Example
smua.source.sink = smua.ENABLE
Also see
Source or sink
2651A-901-01 Rev. A / March 2011
Affected by
SMU reset
Instrument reset
Recall
Set to the source settling mode. Set to one of the following values:
0 or smuX.ENABLE: Turns on sink mode.
1 or smuX.DISABLE: Turns off sink mode.
SMU channel (for example, smua.source.sink applies to SMU Channel A).
(on page 4-5)
Where saved
Default value
Saved setups
smuX.DISABLE