Keithley 2651A Reference Manual page 358

High power system sourcemeter instrument
Hide thumbs Also See for 2651A:
Table of Contents

Advertisement

Section 7: Command reference
Details
Set this attribute to zero (0) to disable the automatic trigger output.
Do not use the stimulus attribute for generating output triggers under script control. Use
digio.trigger[N].assert() instead.
Trigger stimulus for a digital I/O line may be set to one of the existing trigger event IDs, described in the following
table.
Trigger event IDs*
Event ID
smuX.trigger.SWEEPING_EVENT_ID
smuX.trigger.ARMED_EVENT_ID
smuX.trigger.SOURCE_COMPLETE_EVENT_ID
smuX.trigger.MEASURE_COMPLETE_EVENT_ID
smuX.trigger.PULSE_COMPLETE_EVENT_ID
smuX.trigger.SWEEP_COMPLETE_EVENT_ID
smuX.trigger.IDLE_EVENT_ID
digio.trigger[N].EVENT_ID
tsplink.trigger[N].EVENT_ID
lan.trigger[N].EVENT_ID
display.trigger.EVENT_ID
trigger.EVENT_ID
trigger.blender[N].EVENT_ID
trigger.timer[N].EVENT_ID
* Use the name of the trigger event ID to set the stimulus value rather than the numeric value. Using the name
makes the code compatible for future upgrades (for example, if the numeric values must change when
enhancements are added to the instrument).
Example 1
digio.trigger[3].stimulus = 0
Example 2
digio.trigger[3].stimulus =
smua.trigger.SOURCE_COMPLETE_EVENT_ID
7-50
Model 2651A High Power System SourceMeter® Instrument Reference Manual
Event description
Occurs when the SMU transitions from the idle state
to the arm layer of the trigger model
Occurs when the SMU moves from the arm layer to
the trigger layer of the trigger model
Occurs when the SMU completes a source action
Occurs when the SMU completes a measure action
Occurs when the SMU completes a pulse
Occurs when the SMU completes a sweep
Occurs when the SMU returns to the idle state
Occurs when an edge is detected on a digital I/O line
Occurs when an edge is detected on a TSP-Link line
Occurs when the appropriate LXI trigger packet is
received on LAN trigger object N
Occurs when the TRIG key on the front panel is
pressed.
Occurs when a *TRG command is received on the
remote interface.
(GPIB only) Occurs when a GET bus command is
received.
(VXI-11 only) Occurs with the VXI-11 command
device_trigger. Reference the VXI-11 standard
for additional details on the device trigger operation.
Occurs after a collection of events is detected.
Occurs when a delay expires.
Clear the trigger
stimulus of digital
I/O line 3.
Set the trigger
stimulus of digital
I/O line 3 to be the
source complete
event.
2651A-901-01 Rev. A / March 2011