Model 2651A High Power System SourceMeter® Instrument Reference Manual
Syntax rules for instrument commands
Syntax rule
White space: Not
required in a
function.
Function
parameters: All
functions are
required to have a
set of parentheses ()
immediately
following the
function.
Multiple
parameters: Must
be separated by
commas (,).

Logical instruments

You would normally refer to all instrumentation in one enclosure or node as a single instrument. In the
context of Test Script Processor (TSP
think of each individual subdivision in an enclosure, such as a card slot or the channels, as a
stand-alone instrument. To avoid confusion, all subdivisions of the instrumentation in an enclosure
are referred to as "logical instruments."
Each logical instrument is given a unique identifier in a system. These identifiers are used as part of
all commands that control a given logical instrument. The logical instruments are:
beeper
bit
digio
display
eventlog
errorqueue
format
Do not create variable names that are the same as names of logical instruments. Doing so will result
in the loss of use of the logical instrument and its associated commands. For example, if you send the
command digio = 5, you cannot access the digio.* commands until you turn off the power to
the instrument, and then turn it on again.
2651A-901-01 Rev. A / March 2011
Details
Functions can be sent
with or without white
spaces.
You can specify the
function parameters by
placing them between
the parentheses. Note
that the parentheses are
required even when
there are no parameters
specified.
Some commands require
multiple parameters,
which must be separated
by commas (,).
®
) scripting engine and instrument commands, it is useful to
gpib
lan
status
timer
trigger
tsplink
Examples
The following functions, which set digital I/O line
3 low, are equivalent:
digio.writebit(3,0)
digio.writebit (3, 0)
The following function specifies all overlapped
commands in the nodes in group G must
complete before commands from other groups
can execute:
waitcomplete(G)
The command below reads the value of the
local time zone (no parameters are needed):
timezone = localnode.gettimezone()
This command sets the beeper to emit a
double-beep at 2400 Hz, with a beep sequence
of 0.5 seconds on, 0.25 seconds off, and then
0.5 seconds on:
beeper.beep(0.5, 2400)
delay(0.250)
beeper.beep(0.5, 2400)
NOTE
Section 7: Command reference
7-3