Smux.measure.highcrangedelayfactor; Smux.measure.interval - Keithley 2651A Reference Manual

High power system sourcemeter instrument
Hide thumbs Also See for 2651A:
Table of Contents

Advertisement

Model 2651A High Power System SourceMeter® Instrument Reference Manual

smuX.measure.highcrangedelayfactor

This attribute contains a delay multiplier that is used during range changes when High-C mode is active.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
delayFactor = smuX.measure.highcrangedelayfactor
smuX.measure.highcrangedelayfactor = delayFactor
delayFactor
X
Details
This delay multiplier is only active when High-C mode is active.
Example
smua.measure.highcrangedelayfactor = 5
Also see
reset()
(on page 7-151)
setup.recall()
smuX.reset()
smuX.source.highc
High-capacitance mode

smuX.measure.interval

This attribute sets the interval between multiple measurements.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
interval = smuX.measure.interval
smuX.measure.interval = interval
interval
X
2651A-901-01 Rev. A / March 2011
Affected by
Instrument reset
SMU reset
Recall
The delay factor. Set to a value between 1 and 99.
SMU channel (for example, smua.measure.highcrangedelayfactor applies
to SMU Channel A).
(on page 7-170)
(on page 7-209)
(on page 7-215)
(on page 3-65)
Affected by
Instrument reset
SMU reset
Recall
The interval value (in seconds). Set to a value between 0 and 1.
SMU channel (for example, smua.measure.interval applies to SMU
Channel A).
Where saved
Default value
Saved setups
10
Increases the delay for SMU channel A by a factor if 5.
Where saved
Default value
Saved setups
0
Section 7: Command reference
7-199

Advertisement

Table of Contents
loading

Table of Contents