Trigger.blender[N].Stimulus[M] - Keithley 2651A Reference Manual

High power system sourcemeter instrument
Hide thumbs Also See for 2651A:
Table of Contents

Advertisement

Section 7: Command reference
Details
The trigger.blender[N].reset() function resets the following attributes to their factory defaults:
trigger.blender[N].orenable

trigger.blender[N].stimulus[M]

It also clears trigger.blender[N].overrun.
Also see
trigger.blender[N].orenable
trigger.blender[N].overrun
trigger.blender[N].stimulus[M]
trigger.blender[N].stimulus[M]
This attribute specifies which events trigger the blender.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
eventID = trigger.blender[N].stimulus[M]
trigger.blender[N].stimulus[M] = eventID
eventID
N
M
Details
There are four acceptors that can each select a different event. The eventID parameter can be the event ID of
any trigger event.
Use zero to disable the blender input.
The eventID parameter may be one of the existing trigger event IDs shown in the following table.
Trigger event IDs*
Event ID
smuX.trigger.SWEEPING_EVENT_ID
smuX.trigger.ARMED_EVENT_ID
smuX.trigger.SOURCE_COMPLETE_EVENT_ID
smuX.trigger.MEASURE_COMPLETE_EVENT_ID
7-336
Model 2651A High Power System SourceMeter® Instrument Reference Manual
(on page 7-334)
(on page 7-335)
(on page 7-336)
Affected by
Instrument reset
Trigger blender N reset
The event that triggers the blender action; see Details
An integer representing the trigger event blender (1 to 4)
An integer representing the stimulus index (1 to 4)
Where saved
Default value
Not saved
0
Event description
Occurs when the SMU transitions from the idle state
to the arm layer of the trigger model
Occurs when the SMU moves from the arm layer to
the trigger layer of the trigger model
Occurs when the SMU completes a source action
Occurs when the SMU completes a measure action