Model 2651A High Power System SourceMeter® Instrument Reference Manual
Details
Reading this attribute gives the output function of the source for the specified SMU Channel. Setting this attribute
configures the specified SMU Channel as either a voltage source or a current source.
Example
smua.source.func = smua.OUTPUT_DCAMPS
Also see
smuX.source.levelY
smuX.source.output

smuX.source.highc

This attribute sets high capacitance mode.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
highC = smuX.source.highc
smuX.source.highc = highC
highC
X
Details
Turning on High-C mode has the following effects on the SMU settings:
smuX.measure.autorangei is set to smuX.AUTORANGE_FOLLOW_LIMIT and cannot be changed.
Current ranges below 1 μA are not accessible
If smuX.source.limiti is less than 1 μA, it is raised to 1 μA.
If smuX.source.rangei is less than 1 μA, it is raised to 1 μA.
If smuX.source.lowrangei is less than 1 μA, it is raised to 1 μA.
If smuX.measure.lowrangei is less than 1 μA, it is raised to 1 μA.
Example
smua.source.highc = smua.ENABLE
Also see
High-capacitance mode
2651A-901-01 Rev. A / March 2011
(on page 7-216)
(on page 7-221)
Affected by
SMU reset
Instrument reset
Recall
The state of the high capacitance mode. Set to one of the following values:
0 or smuX.DISABLE: Disables high capacitance mode.
1 or smuX.ENABLE: Enables high capacitance mode.
SMU channel (for example, smua.source.highc applies to SMU Channel A).
(on page 3-65)
Selects the source current function
for SMU Channel A.
Where saved
Default value
Saved setups
smuX.DISABLE
Activates high capacitance mode
for SMU Channel A.
Section 7: Command reference
7-215