Model 2651A High Power System SourceMeter® Instrument Reference Manual
Example
digio.trigger[4].mode = 2
Also see
digio.trigger[N].reset()
digio.writebit()
digio.writeport()
Sweep operation
a

digio.trigger[N].overrun

Use this attribute to read the event detector overrun status.
Type
TSP-Link accessible
Attribute (R)
Yes
Usage
overrun = digio.trigger[N].overrun
overrun
N
Details
If this is true, an event was ignored because the event detector was already in the detected state when the
event occurred.
This is an indication of the state of the event detector built into the line itself. It does not indicate if an overrun
occurred in any other part of the trigger model or in any other detector that is monitoring the event.
Example
overrun = digio.trigger[1].overrun
print(overrun)
Also see
digio.trigger[N].reset()
2651A-901-01 Rev. A / March 2011
(on page 7-49)
(on page 7-51)
(on page 7-52)
(on page 3-19)
Affected by
Instrument reset
Digital I/O trigger N clear
Digital I/O trigger N reset
Trigger overrun state (true or false)
Digital I/O trigger line (1 to 14)
(on page 7-49)
Sets the trigger mode for I/O Line 4 to
digio.TRIG_RISING.
Where saved
Default value
Not saved
Not applicable
If there is no trigger overrun, the following
text is output:
false
Section 7: Command reference