Model 2651A High Power System SourceMeter® Instrument Reference Manual
step = (stop - start)/(points -1)
bias level
Example
ConfigPulseIMeasureVSweepLin(smua, 0,
0.01, 0.05, 1, 1e-3, 0.1, 20,
smua.nvbuffer2, 3)
Also see
InitiatePulseTest()

ConfigPulseIMeasureVSweepLog()

This
KIPulse factory script
each point.
Type
TSP-Link accessible
Function
No
2651A-901-01 Rev. A / March 2011
Figure 99: ConfigPulseIMeasureVSweepLin()
step
start
Measurement (1/NPLC seconds)
Pulse initiated here
Last source value before pulse train was initiated
(on page 7-97)
(on page 5-23) function configures a voltage pulse train with a current measurement at
Affected by
step
ton
toff
Set up a pulsed sweep that will use SourceMeter
instrument channel A. The pulsed sweep will start at
10 mA, end at 50 mA, and return to a 0 mA bias level
between pulses. Each pulsed step will be on for 1 ms,
and then at the bias level for 100 ms. The voltage
limit will be 1 V during the entire pulsed sweep. The
pulse train will be comprised of 20 pulsed steps and
the pulse train will be assigned a tag index of 3.
Where saved
Default value
Section 7: Command reference
stop
7-29