Status.operation.instrument - Keithley 2651A Reference Manual

High power system sourcemeter instrument
Hide thumbs Also See for 2651A:
Table of Contents

Advertisement

Model 2651A High Power System SourceMeter® Instrument Reference Manual
As an example, to set bit B1 of the operation status calibration summary enable register, set
status.operation.calibrating.enable = status.operation.calibrating.SMUA.
In addition to the above constant, operationRegister can be set to the decimal equivalent of the bit to set.
Example
status.operation.calibrating.enable =
status.operation.calibrating.SMUA
Also see
Operation Status Registers
status.operation.*

status.operation.instrument.*

This attribute contains the operation status instrument summary register set.
Type
TSP-Link accessible
Attribute
- -
.condition (R)
Yes
.enable (RW)
Yes
.event (R)
Yes
.ntr (RW)
Yes
.ptr (RW)
Yes
Usage
operationRegister = status.operation.instrument.condition
operationRegister = status.operation.instrument.enable
operationRegister = status.operation.instrument.event
operationRegister = status.operation.instrument.ntr
operationRegister = status.operation.instrument.ptr
status.operation.instrument.enable = operationRegister
status.operation.instrument.ntr = operationRegister
status.operation.instrument.ptr = operationRegister
operationRegister
Details
These attributes are used to read or write to the operation status instrument summary registers. Reading a status
register returns a value. The binary equivalent of the returned value indicates which register bits are set. The
least significant bit of the binary number is Bit B0, and the most significant bit is Bit B15. For example, if a value
of 1.02600e+03 (which is 1026) is read as the value of the condition register, the binary equivalent is
0000 0100 0000 0010. This value indicates that bit B1 and bit B10 are set.
2651A-901-01 Rev. A / March 2011
(on page E-20)
(on page 7-266)
Affected by
- -
Not applicable
Status reset
Status reset
Status reset
Status reset
The operation event register's status. A zero (0) indicates no bits set (also send 0 to
clear all bits); other values indicate various bit settings
Sets the SMUA bit of the operation
status calibration summary enable
register using a constant.
Where saved
Default value
- -
- -
Not saved
Not applicable
Not saved
0
Not saved
0
Not saved
0
Not saved
31,746 (All bits set)
Section 7: Command reference
7-269

Advertisement

Table of Contents
loading