Trigger.timer[N].Delaylist; Trigger.timer[N].Event_Id - Keithley 2651A Reference Manual

High power system sourcemeter instrument
Hide thumbs Also See for 2651A:
Table of Contents

Advertisement

Model 2651A High Power System SourceMeter® Instrument Reference Manual
Also see
trigger.timer[N].reset()

trigger.timer[N].delaylist

This attribute sets an array of timer intervals that are used when triggered.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
intervals = trigger.timer[N].delaylist
trigger.timer[N].delaylist = intervals
intervals
N
Details
Each time the timer is triggered, it uses the next delay period from the array.
After all elements in the array have been used, the delays restart at the beginning of the list.
If the array contains more than one element, the average of the delay intervals in the list must be ≥ 50 µs.
Example
trigger.timer[3].delaylist = {50e-6, 100e-6, 150e-6}
DelayList = trigger.timer[3].delaylist
for x = 1, table.getn(DelayList) do
print(DelayList[x])
end
Also see
trigger.timer[N].reset()

trigger.timer[N].EVENT_ID

This constant specifies the trigger timer event number.
Type
TSP-Link accessible
Constant
Yes
2651A-901-01 Rev. A / March 2011
(on page 7-343)
Affected by
Instrument reset
Trigger timer N reset
Table of delay intervals in seconds
Trigger timer number (1 to 4)
(on page 7-343)
Affected by
Where saved
Default value
Not saved
{10e-6}
Set a delay list on trigger timer 3
with three delays (50 µs, 100 µs,
and 150 µs).
Read the delay list on trigger timer
3:
Output (assuming the list was set
with three delays (50 µs, 100 µs,
and 150 µs):
5.000000000e-05
1.000000000e-04
1.500000000e-04
Where saved
Default value
Section 7: Command reference