Section 7: Command reference
Example
SweepILogMeasureV(smua, 0.01, 0.1,
0.001, 5)
Also see
None
SweepVLinMeasureI()
This
KISweep factory script
every step (point).
Type
TSP-Link accessible
Function
No
Usage
SweepVLinMeasureI(smu, startv, stopv, stime, points)
smu
startv
stopv
stime
points
7-328
Model 2651A High Power System SourceMeter® Instrument Reference Manual
(on page 5-22) function performs a linear voltage sweep with current measured at
Affected by
System SourceMeter
Sweep start voltage in volts
Sweep stop voltage in volts
Settling time in seconds; occurs after stepping the source and before performing a
measurement
Number of sweep points (must be ≥2)
This function performs a five-point linear current sweep
starting at 10 mA and stopping at 100 mA. Voltage is
measured at every step (point) in the sweep. The source
will be allowed to settle on each step for 1 ms before a
measurement is performed.
The following table contains log values and corresponding
source levels for the five-point logarithmic sweep:
MeasurePoint
LogStepSize
1
0.25
2
0.25
3
0.25
4
0.25
5
0.25
Where saved
®
instrument channel (set to smua)
LogStep SourceStepLevel
0.0
0.01 A
0.25
0.017783 A
0.5
0.031623 A
0.75
0.056234 A
1.0
0.1 A
Default value
2651A-901-01 Rev. A / March 2011