Smux.measure.filter.enable - Keithley 2651A Reference Manual

High power system sourcemeter instrument
Hide thumbs Also See for 2651A:
Table of Contents

Advertisement

Model 2651A High Power System SourceMeter® Instrument Reference Manual
Details
This attribute sets the size of the stack used for filtered measurements.
Example
smua.measure.filter.count = 10
Also see
reset()
(on page 7-151)
setup.recall()

smuX.measure.filter.enable

smuX.measure.filter.type
smuX.reset()
Filters
(on page 3-3)
smuX.measure.filter.enable
This attribute enables or disables filtered measurements.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
filterState = smuX.measure.filter.enable
smuX.measure.filter.enable = filterState
filterState
X
Details
This attribute enables or disables the filter.
Example
smua.measure.filter.enable = 1
Also see
reset()
(on page 7-151)
setup.recall()
smuX.measure.filter.count
smuX.measure.filter.type
smuX.reset()
Filters
(on page 3-3)
2651A-901-01 Rev. A / March 2011
(on page 7-170)
(on page 7-197)
(on page 7-198)
(on page 7-209)
Affected by
Instrument reset
SMU reset
Recall
The filter status. Set to one of the following values:
0 or smuX.FILTER_OFF: Disables the filter
1 or smuX.FILTER_ON: Enables the filter
SMU channel (for example, smua.measure.filter.enable applies to SMU
Channel A).
(on page 7-170)
(on page 7-196)
(on page 7-198)
(on page 7-209)
Sets the filter count for SMU Channel A to 10.
Where saved
Default value
Saved setups
smuX.FILTER_OFF
Enables the filter for SMU Channel A. Alternatively, the
value 1 may be replaced with smua.FILTER_ON.
Section 7: Command reference
7-197