Status.questionable.over_Temperature - Keithley 2651A Reference Manual

High power system sourcemeter instrument
Hide thumbs Also See for 2651A:
Table of Contents

Advertisement

Section 7: Command reference

status.questionable.over_temperature.*

This attribute contains the questionable status over temperature summary register set.
Type
TSP-Link accessible
Attribute
- -
.condition (R)
Yes
.enable (RW)
Yes
.event (R)
Yes
.ntr (RW)
Yes
.ptr (RW)
Yes
Usage
questionableRegister = status.questionable.over_temperature.condition
questionableRegister = status.questionable.over_temperature.enable
questionableRegister = status.questionable.over_temperature.event
questionableRegister = status.questionable.over_temperature.ntr
questionableRegister = status.questionable.over_temperature.ptr
status.questionable.over_temperature.enable = questionableRegister
status.questionable.over_temperature.ntr = questionableRegister
status.questionable.over_temperature.ptr = questionableRegister
operationRegister
Details
These attributes are used to read or write to the questionable status over temperature summary registers.
Reading a status register returns a value. The binary equivalent of the returned value indicates which register
bits are set. The least significant bit of the binary number is Bit B0, and the most significant bit is Bit B15.
For information about .condition, .enable, .event, .ntr, and .ptr registers, refer to
2) and
Enable and transition registers
meanings:
Bit
Value
B0
Not used
B1
status.questionable.over_temperature.SMUA
B2-B15
Not used
In addition to the above constants, questionableRegister can be set to the numeric equivalent of the bit to
set.
Example
status.questionable.over_temperature.enable =
status.questionable.over_temperature.SMUA
7-306
Model 2651A High Power System SourceMeter® Instrument Reference Manual
Affected by
- -
Not applicable
Status reset
Status reset
Status reset
Status reset
The questionable status over temperature summary register's status. A zero (0)
indicates no bits set (also send 0 to clear all bits); the only valid value other than
zero (0) is 2
(on page E-18). The individual bits of this register have the following
Where saved
Default value
- -
- -
Not saved
Not applicable
Not saved
0
Not saved
0
Not saved
0
Not saved
2 (All bits set)
Status register sets
Description
Not applicable.
Set bit indicates that an over temperature
indication was detected.
Bit B1 decimal value: 2
Binary value: 0000 0010
Not applicable.
Sets the SMU A bit in the
questionable status over
temperature summary enable
register using a constant.
2651A-901-01 Rev. A / March 2011
(on page E-

Advertisement

Table of Contents
loading

Table of Contents