Section 7: Command reference

status.measurement.reading_overflow.*

This attribute contains the measurement event reading overflow summary register set.
Type
TSP-Link accessible
Attribute
- -
.condition (R)
Yes
.enable (RW)
Yes
.event (R)
Yes
.ntr (RW)
Yes
.ptr (RW)
Yes
Usage
measurementRegister = status.measurement.reading_overflow.condition
measurementRegister = status.measurement.reading_overflow.enable
measurementRegister = status.measurement.reading_overflow.event
measurementRegister = status.measurement.reading_overflow.ntr
measurementRegister = status.measurement.reading_overflow.ptr
status.measurement.reading_overflow.enable = measurementRegister
status.measurement.reading_overflow.ntr = measurementRegister
status.measurement.reading_overflow.ptr = measurementRegister
measurementRegister
Details
These attributes are used to read or write to the measurement event reading overflow summary registers.
Reading a status register returns a value. The binary equivalent of the returned value indicates which register
bits are set. The least significant bit of the binary number is Bit B0, and the most significant bit is Bit B15. For
example, assume the value 2 is returned for the enable register. The binary equivalent is 0000 0000 0000 0010.
This value indicates that bit B1 (SMUA) is set.
For information about .condition, .enable, .event, .ntr, and .ptr registers, refer to
2) and
Enable and transition registers
meanings:
7-260
Model 2651A High Power System SourceMeter® Instrument Reference Manual
Affected by
Where saved
- -
- -
Not applicable
Not saved
Status reset
Not saved
Status reset
Not saved
Status reset
Not saved
Status reset
Not saved
The measurement reading overflow summary register status. A zero (0)
indicates no bits set (also send 0 to clear all bits); the only valid value other
than 0 is 2
(on page E-18). The individual bits of this register have the following
Default value
- -
Not applicable
0
0
0
2 (All bits set)
Status register sets
(on page E-
2651A-901-01 Rev. A / March 2011

Advertisement

Table of Contents
loading

Table of Contents