Model 2651A High Power System SourceMeter® Instrument Reference Manual

smuX.trigger.source.action

This attribute enables or disables sweeping the source (on or off).
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
action = smuX.trigger.source.action
smuX.trigger.source.action = action
action
X
Details
This attribute is used to enable or disable source level changes during a sweep. In addition to enabling the action
before initiating the sweep, make sure to configure it using smuX.trigger.source.linearY(),
smuX.trigger.source.listY(), or smuX.trigger.source.logY().
Example
smua.trigger.source.listv({3, 1, 4, 5, 2})
smua.trigger.source.action = smua.ENABLE
Also see
smuX.trigger.source.linearY()
smuX.trigger.source.listY()
smuX.trigger.source.logY()
Triggering
(on page 3-32)
2651A-901-01 Rev. A / March 2011
Affected by
SMU reset
Instrument reset
Sweep source action. Set to one of the following values:
0 or smuX.DISABLE: Do not sweep the source.
1 or smuX.ENABLE: Sweep the source.
SMU channel (for example, smua.trigger.source.action applies to SMU
channel A).
(on page 7-243)
(on page 7-244)
(on page 7-245)
Where saved
Default value
Not saved
smuX.DISABLE