Section 7: Command reference
Details
This function sets the blender event detector to the undetected state and resets the event detector's overrun
indicator.
Also see
None

trigger.blender[N].EVENT_ID

This constant contains the trigger blender event number.
Type
TSP-Link accessible
Constant
Yes
Usage
eventID = trigger.blender[N].EVENT_ID
eventID
N
Details
Set the stimulus of any trigger event detector to the value of this constant to have it respond to trigger events
from this trigger blender.
Also see
None

trigger.blender[N].orenable

This attribute selects whether the blender operates in OR mode or AND mode.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
orenable = trigger.blender[N].orenable
trigger.blender[N].orenable = orenable
orenable
N
Details
This attribute selects whether the blender waits for any one event (the "OR" mode) or waits for all selected
events (the "AND" mode) before signaling an output event.
7-334
Model 2651A High Power System SourceMeter® Instrument Reference Manual
Affected by
Trigger event number
The blender number (1 to 4)
Affected by
Instrument reset
Trigger blender N reset
The orenable mode:
true: OR mode
false: AND mode
The trigger blender (1 to 4)
Where saved
Default value
Where saved
Default value
Not saved
false (AND mode)
2651A-901-01 Rev. A / March 2011