Section 7: Command reference
Bit
Binary value
Decimal
Weights
Example
operationRegister =
status.operation.trigger_overrun.SMUA +
status.operation.trigger_overrun.TRGBLND
status.operation.trigger_overrun.enable = operationRegister
Also see
Operation Status Registers
status.operation.*

status.operation.user.*

These attributes manage the status model's operation status user register set.
Type
TSP-Link accessible
Attribute
- -
.condition (RW)
Yes
.enable (RW)
Yes
.event (R)
Yes
.ntr (RW)
Yes
.ptr (RW)
Yes
Usage
operationRegister = status.operation.user.condition
operationRegister = status.operation.user.enable
operationRegister = status.operation.user.event
operationRegister = status.operation.user.ntr
operationRegister = status.operation.user.ptr
status.operation.user.condition = operationRegister
status.operation.user.enable = operationRegister
status.operation.user.ntr = operationRegister
status.operation.user.ptr = operationRegister
operationRegister
Details
These attributes are used to read or write to the operation status user registers. Reading a status register returns
a value. The binary equivalent of the value indicates which register bits are set. In the binary equivalent, the least
significant bit is bit B0, and the most significant bit is bit B15. For example, if a value of 1.29000e+02 (which is
129) is read as the value of the condition register, the binary equivalent is 0000 0000 1000 0001. This value
indicates that bits B0 and B7 are set.
7-298
Model 2651A High Power System SourceMeter® Instrument Reference Manual
B15
B14
0/1
0/1
32,768
16,384
15
14
(2
)
(2
)
(on page E-20)
(on page 7-266)
Affected by
- -
Status reset
Status reset
Status reset
Status reset
Status reset
The operation status user register's status. A zero (0) indicates no bits set (also
send 0 to clear all bits); other values indicate various bit settings
B13
B12
B11
0/1
0/1
0/1
8,192
4,096
2,048
13
12
11
(2
)
(2
)
(2
)
Where saved
Default value
- -
- -
Not saved
0
Not saved
0
Not saved
0
Not saved
0
Not saved
32,767 (All bits set)
B10
B9
B8
0/1
0/1
0/1
1024
512
256
10
9
8
(2
)
(2
)
(2
)
Sets bit B1 and bit B10 of
the operation status
trigger overrun summary
enable register using
constants.