Trigger.timer[N].Passthrough; Trigger.timer[N].Reset() - Keithley 2651A Reference Manual

High power system sourcemeter instrument
Hide thumbs Also See for 2651A:
Table of Contents

Advertisement

Model 2651A High Power System SourceMeter® Instrument Reference Manual

trigger.timer[N].passthrough

This attribute enables or disables the timer trigger pass-through mode.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
passthrough = trigger.timer[N].passthrough
trigger.timer[N].passthrough = passthrough
passthrough
N
Details
When enabled, triggers are passed through immediately and initiate the delay. When disabled, a trigger only
initiates a delay.
Example
trigger.timer[1].passthrough = true
Also see

trigger.timer[N].reset()

trigger.timer[N].reset()
This function resets some of the trigger timer settings to their factory defaults.
Type
TSP-Link accessible
Function
Yes
Usage
trigger.timer[N].reset()
N
2651A-901-01 Rev. A / March 2011
Affected by
Instrument reset
Trigger timer N reset
The state of pass-through mode. Set to to one of the following values:
true: Enabled
false: Disabled
Trigger timer number (1 to 4)
(on page 7-343)
Affected by
Trigger timer number (1 to 4)
Where saved
Default value
Not saved
false (disabled)
Enables pass-through mode on trigger timer 1.
Where saved
Default value
Section 7: Command reference
7-343

Advertisement

Table of Contents
loading

Table of Contents