Pulsevmeasurei() - Keithley 2651A Reference Manual

High power system sourcemeter instrument
Hide thumbs Also See for 2651A:
Table of Contents

Advertisement

Section 7: Command reference
Example
PulseIMeasureV(smua, 0.001, 1.0,
20E-3, 40E-3, 10)
Also see
None

PulseVMeasureI()

This
KIPulse factory script
Type
TSP-Link accessible
Function
No
Usage
PulseVMeasureI(smu, bias, level, ton, toff, points)
smu
bias
level
ton
toff
points
7-148
Model 2651A High Power System SourceMeter® Instrument Reference Manual
Figure 104: PulseIMeasureV
bias
toff
(on page 5-23) function performs a specified number of pulse V, measure I cycles.
Affected by
System SourceMeter
Bias level in volts
Pulse level in volts
Pulse on time in seconds
Pulse off time in seconds
Number of pulse-measure cycles
Voltage measurement
level
bias
ton
toff
SMU A will output 1 mA and dwell for 40 ms, output
1 A and dwell for 20 ms. The voltage measurement
occurs during the 20 ms dwell periods. After the
measurement, the output will return to 1 mA and
dwell for 40 ms. This pulse-measure process will
repeat nine more times.
Where saved
®
instrument channel (set to smua)
Default value
2651A-901-01 Rev. A / March 2011