Smux.trigger.endpulse.set(); Smux.trigger.endpulse.stimulus - Keithley 2651A Reference Manual

High power system sourcemeter instrument
Hide thumbs Also See for 2651A:
Table of Contents

Advertisement

Model 2651A High Power System SourceMeter® Instrument Reference Manual
Example
smua.trigger.endpulse.action =
smua.SOURCE_IDLE
smua.trigger.endpulse.stimulus =
trigger.timer[1].EVENT_ID
Also see
Triggering
(on page 3-32)

smuX.trigger.endpulse.set()

This function sets the end pulse event detector to the detected state..
Type
TSP-Link accessible
Function
Yes
Usage
smuX.trigger.endpulse.set()
X
Details
This function sets the end pulse event detector to the detected state.
The SMU will automatically clear all the event detectors when the smuX.trigger.initiate() function is
executed. This function should be called after the sweep is initiated. Make sure that if the event detectors are
configured to automatically clear (because the smuX.trigger.autoclear attribute is set to smuX.ENABLE)
that this command is issued after the SMU has entered the trigger layer.
Also see
smuX.trigger.autoclear
smuX.trigger.initiate()
Triggering
(on page 3-32)

smuX.trigger.endpulse.stimulus

This attribute defines which event will cause the end pulse event detector to enter the detected state.
Type
TSP-Link accessible
Attribute (RW)
Yes
2651A-901-01 Rev. A / March 2011
Affected by
Source-measure unit (SMU) channel (for example,
smua.trigger.endpulse.set() applies to SMU channel A).
(on page 7-229)
(on page 7-234)
Affected by
SMU reset
Instrument reset
Configure end pulse action to achieve a
pulse and configures trigger timer 1 to
control the end of pulse.
Where saved
Default value
Where saved
Default value
Not saved
0