Tsplink.trigger[N].Wait() - Keithley 2651A Reference Manual

High power system sourcemeter instrument
Hide thumbs Also See for 2651A:
Table of Contents

Advertisement

Section 7: Command reference
Trigger event IDs*
Event ID
trigger.EVENT_ID
trigger.blender[N].EVENT_ID
trigger.timer[N].EVENT_ID
* Use the name of the trigger event ID to set the stimulus value rather than the numeric value. Using the name
makes the code compatible for future upgrades (for example, if the numeric values must change when
enhancements are added to the instrument).
Also see
tsplink.trigger[N].assert()
tsplink.trigger[N].reset()

tsplink.trigger[N].wait()

This function waits for a trigger.
Type
TSP-Link accessible
Function
Yes
Usage
triggered = tsplink.trigger[N].wait(timeout)
triggered
N
timeout
Details
This function waits up to the timeout value for an input trigger. If one or more trigger events were detected since
the last time tsplink.trigger[N].wait() or tsplink.trigger[N].clear() was called, this function
returns immediately.
After waiting for a trigger with this function, the event detector is automatically reset and rearmed. This is true
regardless of the number of events detected.
Example
triggered = tsplink.trigger[3].wait(10)
print(triggered)
7-358
Model 2651A High Power System SourceMeter® Instrument Reference Manual
(on page 7-350)
(on page 7-356)
Affected by
Trigger detection indication. Set to one of the following values:
true: A trigger is detected during the timeout period
false: A trigger is not detected during the timeout period
The trigger line (1 to 3)
The timeout value in seconds
Event description
Occurs when a *TRG command is received on the
remote interface.
(GPIB only) Occurs when a GET bus command is
received.
(VXI-11 only) Occurs with the VXI-11 command
device_trigger. Reference the VXI-11 standard
for additional details on the device trigger operation.
Occurs after a collection of events is detected.
Occurs when a delay expires.
Where saved
Default value
Waits up to 10 seconds for a trigger
on TSP-Link line 3.
If false is returned, no trigger was
detected during the 10-second
timeout.
If true is returned, a trigger was
detected.
2651A-901-01 Rev. A / March 2011

Advertisement

Table of Contents
loading

Table of Contents