Model 2651A High Power System SourceMeter® Instrument Reference Manual
Example
ConfigPulseIMeasureV(smua, 0, 5, 10,
0.001, 0.080, 1, smua.nvbuffer1, 1)
Also see
InitiatePulseTest()

ConfigPulseIMeasureVSweepLin()

This
KIPulse factory script
measurement at each point.
Type
TSP-Link accessible
Function
No
Usage
f, msg = ConfigPulseIMeasureVSweepLin(smu, bias, start, stop, limit, ton, toff,
points, buffer, tag, sync_in, sync_out, sync_in_timeout, sync_in_abort)
f, msg = ConfigPulseIMeasureVSweepLin(smu, bias, start, stop, limit, ton, toff,
points, buffer, tag, sync_in, sync_out, sync_in_timeout)
f, msg = ConfigPulseIMeasureVSweepLin(smu, bias, start, stop, limit, ton, toff,
points, buffer, tag, sync_in, sync_out)
f, msg = ConfigPulseIMeasureVSweepLin(smu, bias, start, stop, limit, ton, toff,
points, buffer, tag, sync_in)
f, msg = ConfigPulseIMeasureVSweepLin(smu, bias, start, stop, limit, ton, toff,
points, buffer, tag)
2651A-901-01 Rev. A / March 2011
(on page 7-97)
(on page 5-23) function configures a linear pulsed current sweep with a voltage
Affected by
Set up a pulse train that uses System SourceMeter
instrument channel A. The pulse amplitude will be 5 A
and will return to 0 A after 1 ms. The pulse will remain
at 0 A for 80 ms and the voltage limit will be 10 V
during the pulse. The pulse train will consist of only 1
pulse, and this pulse will be assigned a tag index of
1.