Trigger.timer[N].Stimulus - Keithley 2651A Reference Manual

High power system sourcemeter instrument
Hide thumbs Also See for 2651A:
Table of Contents

Advertisement

Section 7: Command reference
Details
The trigger.timer[N].reset() function resets the following attributes to their factory defaults:
trigger.timer[N].count
trigger.timer[N].delay
trigger.timer[N].delaylist
trigger.timer[N].passthrough

trigger.timer[N].stimulus

It also clears trigger.timer[N].overrun.
Also see
trigger.timer[N].count
trigger.timer[N].delay
trigger.timer[N].delaylist
trigger.timer[N].overrun
trigger.timer[N].passthrough
trigger.timer[N].stimulus
trigger.timer[N].stimulus
This attribute specifies which event starts the timer.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
eventID = trigger.timer[N].stimulus
trigger.timer[N].stimulus = eventID
eventID
N
Details
The eventID parameter may be one of the trigger event IDs shown in the following table.
Trigger event IDs*
Event ID
smuX.trigger.SWEEPING_EVENT_ID
smuX.trigger.ARMED_EVENT_ID
smuX.trigger.SOURCE_COMPLETE_EVENT_ID
smuX.trigger.MEASURE_COMPLETE_EVENT_ID
7-344
Model 2651A High Power System SourceMeter® Instrument Reference Manual
(on page 7-339)
(on page 7-340)
(on page 7-341)
(on page 7-342)
(on page 7-343)
(on page 7-344)
Affected by
Instrument reset
Trigger timer N reset
The event that triggers the timer delay
A number representing a trigger timer number (1 to 4)
Where saved
Default value
Not saved
0
Event description
Occurs when the SMU transitions from the idle state
to the arm layer of the trigger model
Occurs when the SMU moves from the arm layer to
the trigger layer of the trigger model
Occurs when the SMU completes a source action
Occurs when the SMU completes a measure action
2651A-901-01 Rev. A / March 2011

Advertisement

Table of Contents
loading

Table of Contents