Keithley 2651A Reference Manual page 203

High power system sourcemeter instrument
Hide thumbs Also See for 2651A:
Table of Contents

Advertisement

Section 4: Theory of operation
The first graph in the figure, labeled "A: Output characteristics," shows the output characteristics for
the I-source. As shown, Model 2651A instruments can continuously output up to 20.2 A at 10 V, up to
10.1 A at 20 V, or up to 5.05 A at 40 V. Note that when continuously sourcing more than 5.05 A,
voltage is limited to 20 V, and when sourcing more than 10.1 A, voltage is limited to 10 V.
The second graph in the figure, labeled "B: Limit lines," shows the limit lines for the I-source. The
current source limit line represents the maximum source value possible for the presently selected
current source range. The voltage compliance limit line represents the actual compliance that is in
effect (see
Compliance principles
operating limits of the System SourceMeter instrument for this quadrant of operation. The operating
point can be anywhere inside (or on) these limit lines. The limit line boundaries for the other
quadrants are similar.
4-6
Model 2651A High Power System SourceMeter® Instrument Reference Manual
Figure 71: Models 2651A I-source boundaries
Limit V
40 V
20 V
10 V
5.05 A
10.1 A
A: Output characteristics
Voltage
compliance
limit line
V-measure
I-source
B: Limit lines
(on page 4-2)). These limit lines are boundaries that represent the
Source I
20.2 A
Current source
limit line
2651A-901-01 Rev. A / March 2011

Advertisement

Table of Contents
loading

Table of Contents