Section 5: Remote commands

Userstrings

Use the functions in this group to store and retrieve user-defined strings in nonvolatile memory. These
strings are stored as key-value pairs. You can use the userstring functions to store custom,
instrument-specific information in the instrument, such as department number, asset number, or
manufacturing plant location.
userstring.add()
userstring.catalog()
userstring.delete()
userstring.get()

Factory scripts

Introduction

The Keithley Instruments Model 2651A High Power System SourceMeter
one or more factory scripts saved in its flash firmware memory. A factory script is made up of a
number of functions. Some of them can be called from the front-panel LOAD TEST menu. All of them
can be called using remote programming.
As Keithley Instruments develops additional factory scripts, they will be made available on the
Keithley Instruments website
2651A. See
your Model 2651A.
A factory script is similar to a user script, except a factory script is created by Keithley Instruments at
the factory and is permanently stored in nonvolatile memory. The differences between a user script
and a factory script include the following:
A factory script cannot be deleted from nonvolatile memory.
The script listing for a factory script can be retrieved and modified, but it will then be treated as a
user script. A user script cannot be saved as a factory script.
Factory scripts are not stored in global variables. The only references to factory scripts are in the
script.factory.scripts attribute.
The script.factory.catalog() function returns an iterator that can be used in a for loop
to iterate over all the factory scripts.
Example
To retrieve the catalog listing for factory scripts, send:
for name in script.factory.catalog() do print(name) end
5-20
Model 2651A High Power System SourceMeter® Instrument Reference Manual
(on page 7-372)
(on page 7-372)
(on page 7-373)
(on page 7-374)
(http://www.keithley.com) as a flash firmware upgrade for the Model
Upgrading the firmware
(on page A-4) for instructions on upgrading the flash firmware of
®
Instrument is shipped with