smuX.measure.count

This attribute sets the number of measurements performed when a measurement is requested.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
count = smuX.measure.count
smuX.measure.count = count
count
X
Details
This attribute controls the number of measurements taken any time a measurement is requested. When using a
reading buffer with a measure command, this attribute also controls the number of readings to be stored.
If smuX.measure.count is set to a value greater than 1, any measurement delay set by
smuX.measure.delay will only occur before the first measurement, while the smuX.measure.interval
controls the interval between successive measurements.
Example
smua.measure.count = 10
Also see
reset()
(on page 7-151)
smuX.measure.delay
smuX.measure.interval
smuX.measure.overlappedY()
smuX.measure.Y()
smuX.reset()
setup.recall()
7-194
Model 2651A High Power System SourceMeter® Instrument Reference Manual
(on page 7-177)
(on page 7-179)
(on page 7-180)
(on page 7-181)
(on page 7-211)
Affected by
Instrument reset
SMU reset
Recall
Number of measurements.
SMU channel (for example, smua.measure.count applies to SMU Channel A).
(on page 7-195)
(on page 7-199)
(on page 7-201)
(on page 7-205)
(on page 7-209)
(on page 7-170)
Where saved
Default value
Saved setups
1
Sets the SMU Channel A measure count to 10.