Smux.trigger.measure.set() - Keithley 2651A Reference Manual

High power system sourcemeter instrument
Hide thumbs Also See for 2651A:
Table of Contents

Advertisement

Section 7: Command reference
Details
With this attribute enabled (setting action to smuX.ENABLE or smuX.ASYNC), configure the measurement with
one of the smuX.trigger.measure.Y() functions.
If this attribute is set to smuX.ASYNC:
Asynchronous sweep measurements can only be used with measure autoranging turned off. To turn
measure autoranging off for all measurements during the sweep, set the
smuX.measure.autorangeY attribute to smuX.AUTORANGE_OFF.
If the integrating ADC is selected (smuX.measure.adc attribute is set to smuX.ADC_INTEGRATE),
autozero must also be turned off. To turn off autozero, set the smuX.measure.autozero attribute to
smuX.AUTOZERO_OFF or smuX.AUTOZERO_ONCE.
If either of the above items are incorrectly configured, the smuX.trigger.initiate() function will generate
an error.
Also see
smuX.measure.adc
smuX.trigger.autoclear
smuX.trigger.measure.Y()
Triggering
(on page 3-32)

smuX.trigger.measure.set()

This function sets the measure event detector to the detected state.
Type
TSP-Link accessible
Function
Yes
Usage
smuX.trigger.measure.set()
X
Details
This function is useful whenever you want the SMU to continue operation without waiting for a programmed
trigger event. When called, this function immediately satisfies the event detector, allowing the SMU to continue
through the trigger model.
A common example of when this function can be used is when you want the SMU to immediately perform an
action the first time through the trigger model even if a programmed trigger event does not occur. Make sure that
if the event detectors are configured to automatically clear (because the smuX.trigger.autoclear attribute
is set to smuX.ENABLE) that this command is issued after the SMU has entered the trigger layer. This function
can also be used to start actions on the SMU in case of a missed trigger event.
The SMU will automatically clear all event detectors when the smuX.trigger.initiate() function is
executed. This function should be called after the sweep is initiated.
Example
smua.trigger.measure.set()
7-236
Model 2651A High Power System SourceMeter® Instrument Reference Manual
(on page 7-190)
(on page 7-229)
(on page 7-238)
Affected by
Source-measure unit (SMU) channel (for example,
smua.trigger.measure.set() applies to SMU channel A)
Where saved
Default value
Sets the measure event detector of
SMU channel A.
2651A-901-01 Rev. A / March 2011

Advertisement

Table of Contents
loading

Table of Contents