These attributes are used to read or write to the standard event status registers. Reading a status register
returns a value. The binary equivalent of the returned value indicates which register bits are set. The least
significant bit of the binary number is Bit B0, and the most significant bit is Bit B15. For example, if a value of
1.29000e+02 (which is 129) is read as the value of the condition register, the binary equivalent is
0000 0000 1000 0001. This value indicates that bit B0 and bit B7 are set.
B15
B14
**
>
0
0
* Least significant bit
** Most significant bit
For information about .condition, .enable, .event, .ntr, and .ptr registers, refer to
2) and
Enable and transition registers
meanings:
Bit
Value
B0
status.standard.OPERATION_COMPLETE
status.standard.OPC
B1
Not used
B2
status.standard.QUERY_ERROR
status.standard.QYE
B3
status.standard.DEVICE_DEPENDENT_ERROR
status.standard.DDE
B4
status.standard.EXECUTION_ERROR
status.standard.EXE
B5
status.standard.COMMAND_ERROR
status.standard.CME
B6
status.standard.USER_REQUEST
status.standard.URQ
2651A-901-01 Rev. A / March 2011
B13
B12
B11
B10
>
>
>
>
0
0
0
0
(on page E-18). The individual bits of this register have the following
B9
B8
B7
B6
>
>
>
>
0
0
1
0
Section 7: Command reference
B5
B4
B3
B2
>
>
>
>
0
0
0
0
Status register sets
Description
Set bit indicates that all pending selected
instrument operations are completed and the
instrument is ready to accept new commands.
The bit is set in response to an *OPC
command. The opc() function can be used in
place of the *OPC command.
Bit B0 decimal value: 1
Not applicable
Set bit indicates that you attempted to read
data from an empty Output Queue.
Bit B2 decimal value: 4
Set bit indicates that an instrument operation
did not execute properly due to some internal
condition.
Bit B3 decimal value: 8
Set bit indicates that the instrument detected an
error while trying to execute a command.
Bit B4 decimal value: 16
Set bit indicates that a command error has
occurred. Command errors include:
IEEE Std 488.2 syntax error: Instrument
received a message that does not follow the
defined syntax of the IEEE Std 488.2 standard.
Semantic error: Instrument received a
command that was misspelled or received an
optional IEEE Std 488.2 command that is not
implemented.
GET error: The instrument received a Group
Execute Trigger (GET) inside a program
message.
Bit B5 decimal value: 32
Set bit indicates that the LOCAL key on the
instrument front panel was pressed.
Bit B6 decimal value: 64
B1
B0
>
*
0
1
(on page E-
7-313