Smux.source.limity - Keithley 2651A Reference Manual

High power system sourcemeter instrument
Hide thumbs Also See for 2651A:
Table of Contents

Advertisement

Model 2651A High Power System SourceMeter® Instrument Reference Manual

smuX.source.limitY

This attribute sets compliance limits.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
limit = smuX.source.limitY
smuX.source.limitY = limit
limit
X
Y
Details
Use the smuX.source.limiti attribute to limit the current output of the voltage source. Use
smuX.source.limitv to limit the voltage output of the current source. The SMU will always choose
(autorange) the source range for the limit setting. Use the smuX.source.limitp attribute to limit the output
power of the source. This attribute should be set in the test sequence before the turning the source on.
Using a limit value of 0 will result in a "Parameter Too Small" error message (error 1102) for v and i. Setting
this attribute to zero disables power compliance for p. When setting the power compliance limit to a nonzero
value, the SMU will adjust the source limit where appropriate to limit the output to the specified power. The SMU
will use the lower of the programmed compliance value (the compliance level that would be used if power
compliance were disabled) or the limit calculated from the power compliance setting.
Reading this attribute indicates the presently set compliance value. Use smuX.source.compliance to read
the state of source compliance.
Example
smua.source.limitv = 30
Also see
smuX.source.compliance
smuX.source.func
smuX.source.output
DUT test connections
2651A-901-01 Rev. A / March 2011
Affected by
SMU reset
Instrument reset
Recall
The compliance limit value. Set to one of the following values: Set voltage
compliance from 0 V to 40 V
Set current compliance from 0 A to 20 A
Set power compliance in watts.
Source-measure unit (SMU) channel (for example, smua.source.limitv applies
to SMU channel A)
SMU function (v = voltage, i = current, p = power)
(on page 7-213)
(on page 7-214)
(on page 7-221)
(on page 2-42)
Where saved
Default value
Saved setups
Limit voltage: 20 (20V)
Limit current: 100e-3 (100 mA)
Limit power: 0 (disabled)
Sets voltage source compliance of
SMU Channel A to 30 V.
Section 7: Command reference