Status.operation.calibrating - Keithley 2651A Reference Manual

High power system sourcemeter instrument
Hide thumbs Also See for 2651A:
Table of Contents

Advertisement

Section 7: Command reference
Example 2
-- decimal 20480 = binary 0101 0000 0000 0000
operationRegister = 20480
status.operation.enable = operationRegister
Also see
Operation Status Registers

status.operation.calibrating.*

This attribute contains the operation status calibration summary register set.
Type
TSP-Link accessible
Attribute
- -
.condition (R)
Yes
.enable (RW)
Yes
.event (R)
Yes
.ntr (RW)
Yes
.ptr (RW)
Yes
Usage
operationRegister = status.operation.calibrating.condition
operationRegister = status.operation.calibrating.enable
operationRegister = status.operation.calibrating.event
operationRegister = status.operation.calibrating.ntr
operationRegister = status.operation.calibrating.ptr
status.operation.calibrating.enable = operationRegister
status.operation.calibrating.ntr = operationRegister
status.operation.calibrating.ptr = operationRegister
operationRegister
Details
These attributes are used to read or write to the operation status calibration summary registers. Reading a status
register returns a value. The binary equivalent of the returned value indicates which register bits are set. The
least significant bit of the binary number is Bit B0, and the most significant bit is Bit B15.
For information about .condition, .enable, .event, .ntr, and .ptr registers, refer to
2) and
Enable and transition registers
meanings:
Bit
Value
B0
Not used
B1
status.operation.calibrating.SMUA
B2 to
Not used
B15
7-268
Model 2651A High Power System SourceMeter® Instrument Reference Manual
(on page E-20)
Affected by
- -
Not applicable
Status reset
Status reset
Status reset
Status reset
The operation calibrating event register's status. A zero (0) indicates no bits set
(also send 0 to clear all bits); the only valid value other than 0 is 2
(on page E-18). The individual bits of this register have the following
Sets the USER and PROG bits of the
operation status enable register using a
decimal value.
Where saved
Default value
- -
- -
Not saved
Not applicable
Not saved
0
Not saved
0
Not saved
0
Not saved
2 (All bits set)
Status register sets
Description
Not applicable.
Set bit indicates that SMU A is unlocked for
calibration.
Bit B1 decimal value: 2
Binary value: 0000 0010
Not applicable.
2651A-901-01 Rev. A / March 2011
(on page E-

Advertisement

Table of Contents
loading

Table of Contents